Changes for page ANU Seismic Data Loggers
Last modified by robert on 2025/04/24 14:58
Summary
-
Page properties (1 modified, 0 added, 0 removed)
Details
- Page properties
-
- Content
-
... ... @@ -18,7 +18,7 @@ 18 18 19 19 The format for "ANUSRSetup.txt" / LPR200s will be a single line of text that looks like this: 20 20 21 -XXX195G0100010034864 21 +{{{XXX195G0100010034864 2 }}} 22 22 23 23 corresponding to site XXX19, 24 24 network 5G, ... ... @@ -26,17 +26,16 @@ 26 26 "gps interval" 01 (once per hour~-~- don't change), 27 27 start mode (always 0), 28 28 stop mode (always 0), 29 - seismometer type (3 = broadband, 1 = shortperiod), 29 + seismometer type (3 = broadband (+/- 20V), 1 = shortperiod (+/- 5V), 30 30 and seismometer serial number (4864). 31 31 the 2 at the very end is for "start recording on power" 32 - and is vital to have this set so the logger will 33 - resume recording on its own if it experiences a power cycle. 32 + and is vital to have this set so the logger will resume recording on its own if it experiences a power cycle. (# of blank spaces before this doesn't matter) 34 34 35 35 the TSAWRs are a bit shorter as they don't have an option to set the "gps sync" 36 36 37 -SITE1XX02500039999 36 +{{{SITE1XX02500039999 2}}} 38 38 39 -which would set up a trillium with serial 9999 named XX.SITE1 at 250hz sample rate.38 +which would set up a Trillium Compact with serial 9999 named XX.SITE1 at 250hz sample rate. 40 40 41 41 42 42 Here is a general guideline for how much data you can expect to fit on a card: