6 (((
7 (% class="container" %)
8 (((
9 ANU has designed and built three generations of seismic data recorders, two of which (the small yellow TerraSAWR and the larger beige LPR-200) are still in use today. Despite their contrasting appearance, they both use the same software, data cards, and have the same instrument response.
10 )))
11 )))
12
13 = Data Card Formatting and Information =
14
15 Both the TerraSAWR and LPR-200 require SD Cards to be formatted in FAT32 filesystem. For 64Gb cards it can be difficult to format in FAT32, but [[software >>http://auspass.edu.au/field/fat32cardformatter.exe]]is available. ANU recommend SanDisk Extreme 150 mb/s cards in either 32 or 64Gb size. We strongly discourage using cards larger than 64Gb, and in general smaller cards are less likely to fail. We have also found that "adapter" cards (e.g. SD to microSD) are prone to having write issues and **strongly** advise against them.
16
17 The loggers can be "pre-programmed" with information (e.g. site name, sampling rate, etc) or they can be programmed in the field using the buttons on the logger. To pre-program the cards you simply edit a text file (named "[[ANUSRSetup.txt>>http://auspass.edu.au/field/ANUSRSetup.txt]]" for the LPRs, or "[[tSAWRSetup.txt>>http://auspass.edu.au/field/tSAWRSetup.txt]]" for the TerraSAWRs) and place it in the root directory on the SDCard. When the logger boots up, it will parse and load this information.
18
19 === The format for ANUSRSetup.txt / LPR200s will be a single line of text that looks like this: ===
20
21 {{{XXX195G0100010034864 2 }}}
22
23 corresponding to site XXX19,
24 network 5G,
25 sample rate 0100, (i.e. 100 Hz)
26 "gps interval" 01 (once per hour~-~- don't change),
27 start mode (always 0),
28 stop mode (always 0),
29 seismometer type (3 = broadband (+/- 20V), 2 = Guralp (+/- 10V), 1 = shortperiod (+/- 5V),
30 and seismometer serial number (4864).
31
32 (% class="box warningmessage" %)
33 (((
34 NOTE: the 2 at the very end is for "RECORD ON RESTART". The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume. (# of blank spaces before this doesn't matter)
35 )))
36
37 === The format for TSAWR loggers is shorter ===
38
39 (% class="box errormessage" %)
40 (((
41 **NOTE: The format is shorter as they don't have an option to set the "GPS sync". The GPS sync number should be omitted otherwise the formatting will not work!**
42 )))
43
44 SITE15G02500039999 2
45
46 corresponding to site SITE1,
47
48 network 5G
49 sample rate 0250, (i.e. 250 Hz)
50 start mode (always 0),
51 stop mode (always 0),
52 seismometer type (3 = broadband (+/- 20V), 2 = Guralp (+/- 10V), 1 = shortperiod (+/- 5V),
53 and seismometer serial number (9999).
54
55
56 === The formatting process using the logger: ===
57
58 The process for formatting an SD card within the logger is straightforward. Navigate to the "SD INFORMATION" screen and press ERASE SD CARD. This process may take up to a minute. This will result in erasing all files from the card. Upon starting recording, a new 'seed' will be written containing all the information that the logger has been set with FINISH THIS SECTION
59
60
61 Here is a general guideline for how much data you can expect to fit on a card:
62
63 {{{ 64Gb card: 245 days @ 250hz or 610 days @ 100hz
64 32Gb card: 122 days @ 250hz or 305 days @ 100hz
65 }}}
66
67
68 = Logger Menus Overview (and setup) =
69
70 == System Information ==
71
72 This menu provides an overview of the status of the station. Under 'Initialisation Parameters', the status of systems that are initialised upon powering on the logger can be checked.
73
74 This menu also displays the firmware version, battery, external, and solar voltages, and the temperature of the system.
75
76 ==== Upon setup ====
77
78 - Check all Initialisation Parameters are marked as successful.
79
80 - Check that solar voltage is above 10 V, otherwise the station will not last long.
81
82
83 == Live Seismometer Data ==
84
85 This screen displays real time seismometer data for all 3 axes of the seismometer. Initially, a plot of raw data from all 3 channels is shown. By pressing enter, a high pass filter can be applied showing a more useful plot of all channels. The up and down arrows can be used to change between a view of all channels, individual channels, and all channels on different X axes.
86
87 ==== Upon setup ====
88
89 - Check that all 3 channels are present and are producing a signal.
90
91 - Stomp on the ground near where the sensor is installed to check that a signal is being picked up by all 3 channels.
92
93 - Common representations of poor signals include channels sloping smoothly from high values to zero, or sharp corners in the plots.
94
95
96 == GPS ==
97
98 This menu displays the status of the stations' GPS connection. The screen lists; UTC time, UTC date, latitude, longitude, altitude, number of satellite connections, and SNR.
99
100 ==== Upon setup ====
101
102 - Check that the station is connected to satellites
103
104
105 == SD Information ==
106
107 The menu lists if the SD card has been initialised, as well as the capacity and space free on the memory card. See "Data card formatting" (above) to see how to format and initialise the inserted SD card.
108
109
110 == System Configuration ==
111
112 This menu is used to set the stations identifiers and parameters.
113
114 First, the sampling rate can be chosen from a set of options (1 Hz, 10 Hz, 25 Hz, 40 Hz, 50 Hz, 100 Hz, 250 Hz, 1000 Hz).
115
116 Next, the station identifier (up to 5 characters) and the network code (2 characters) can be set.
117
118 Record start and stop mode should be left to 'On Request' mode.
119
120 The seismometer model and serial number (up to 10 characters) can be set. Seismometer model options include:
121
122 - Trillium Compact
123
124 - CMG - 3ESP
125
126 - Guralp 40T
127
128 - LE-3D Lite
129
130 - Mark L4C
131
132 - Mark L4
133
134 Additionally, the following settings can be used in place of the seismometer model types: +/- 20V, +/- 10V, +/- 5V
135
136 The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume.
137
138 ==== Upon setup ====
139
140 - Set the sampling rate, station identifier, and network code. Set the appropriate seismometer type. Most importantly, ensure the record on restart setting is checked.
141
142 (% class="box errormessage" %)
143 (((
144 NOTE: Ensure the 'RECORD ON RESTART' option is marked with a cross.
145 )))
146
147 == STA/LTA Configuration ==
148
149 -
150
151 = =
152
153 = LogFile Conversion Script =
154
155 Both the TSAWR and LPR-200 write logging information as a binary "dat" file which includes GPS time syncs, temperatures, battery power, and position. To convert them into ASCII you can read, use [[THIS PYTHON SCRIPT>>http://auspass.edu.au/field/anusr_log.py]].
156
157 = Instrument Response =
158
159 Both the TerraSAWR and LPR-200 use the same ADS1281 analog-to-digital converter chip and are designed to have identical instrument response. Depending on the output sample rate (e.g. 100 Hz, 250 Hz, 500 Hz, 1000 Hz) amplitude response is consistently flat up to ~~100 Hz but phase response can vary above 1 Hz at 100 Hz (or 10 Hz at 250 Hz).
160
161 The user can choose to apply a 2nd stage "sensor gain" by selecting an instrument type in the setup menu. This effectively selects a 10 Vpp (e.g. short period sensors), 20 Vpp, 40 Vpp (most broadband sensors) regime to match the sensor's sensitivity. This has the effect of doubling amplitude from 10v to 20v, or quadrupling from 10v to 40v. If you have set your sensor correctly (and the signal isn't clipped!) you can "correct" this by simply multiplying your data by 0.5 etc.
162
163 Another important thing to note is that the group delay associated with late stage FIR filters is **automatically applied in the logger**, hence there is no need to apply this in the response. These tend to max out at 0.124 seconds for most output sampling rates (0.062 s for 100 Hz).
164
165 Instrument response can be downloaded from IRIS-NRL (v2) if need be, or by downloading the response of an equivalent sensor at AusPass (e.g. get_stations(level='response') )
166
167
168 = ANU TerraSAWR (Gen 3, FW 3.5a, 2017?- current) =
169
170 Text here
171
172 == Sub-paragraph ==
173
174 XXXX
175
176 == Sub-paragraph ==
177
178 xx
179
180 === Sub-sub paragraph ===
181
182 x
183
184
185 = ANU LPR-200 (Gen 2, FW 2.6a/2.7a, 2013 - current) =
186
187 Text here
188
189 == Sub-paragraph ==
190
191 x
192
193 == Sub-paragraph ==
194
195 x
196
197
198 = ANU "ANUSR" (Gen 1, 2003? - 2013?) =
199
200 x
201
202 = References =
203
204 * [[PDF Manual>>http://auspass.edu.au/field/LPR-200_Instruction_Manual.pdf]]
205 )))
206
207 (% class="col-xs-12 col-sm-4" %)
208 (((
209 (% class="box" %)
210 (((
211 **Contents**
212
213 {{toc/}}
214 )))
215
216
217 (% class="box" %)
218 (((
219 = TerraSAWR Specs =
220
221
222 [[image:Terrasawr 1.jpg]]
223
224 [[image:1704865206800-824.jpg]]
225
226 [[image:1704864886951-793.jpg]]
227
228 |=Size (L x W x H)|30 x 25 x12cm
229 |=Weight (with battery)| ~*~*1.9kg (2.5kg)
230 |=Battery|(((
231 Lead-acid 12v, ~~7Ah
232
233 Rechargeable
234 )))
235
236 [[image:terrasawr battery.jpg]]
237 )))
238
239
240 (% class="box" %)
241 (((
242 = LPR-200 Specs =
243
244 [[image:LPR 1.jpg]]
245
246 [[image:LPR 2.jpg]]
247
248 [[image:LPR 3.jpg]]
249
250 |=Size (L x W x H)|42 x 34 x17 cm
251 |=Weight (with battery)|4.9 kg (7.3 kg)
252 |=Battery|(((
253 LiFePo4 140Ah
254
255 Rechargeable
256 )))
257
258 [[image:LPR battery.jpg]]
259 )))
260 )))