6 (((
7 (% class="container" %)
8 (((
9 ANU has designed and built three generations of seismic data recorders, two of which (the small yellow TerraSAWR and the larger beige LPR-200) are still in use today. Despite their contrasting appearance, they both use the same software, data cards, and have the same instrument response.
10 )))
11 )))
12
13 = Data Card Formatting and Information =
14
15 Both the TerraSAWR and LPR-200 require SD Cards to be formatted in FAT32 filesystem. For 64Gb cards it can be difficult to format in FAT32, but [[software >>http://auspass.edu.au/field/fat32cardformatter.exe]]is available. ANU recommend SanDisk Extreme 150 mb/s cards in either 32 or 64Gb size. We strongly discourage using cards larger than 64Gb, and in general smaller cards are less likely to fail. We have also found that "adapter" cards (e.g. SD to microSD) are prone to having write issues and **strongly** advise against them.
16
17 The loggers can be "pre-programmed" with information (e.g. site name, sampling rate, etc) or they can be programmed in the field using the buttons on the logger. To pre-program the cards you simply edit a text file (named "[[ANUSRSetup.txt>>http://auspass.edu.au/field/ANUSRSetup.txt]]" for the LPRs, or "[[tSAWRSetup.txt>>http://auspass.edu.au/field/tSAWRSetup.txt]]" for the TerraSAWRs) and place it in the root directory on the SDCard. When the logger boots up, it will parse and load this information.
18
19 === The format for ANUSRSetup.txt / LPR200s will be a single line of text that looks like this: ===
20
21 {{{XXX195G0100010034864 2 }}}
22
23 corresponding to site XXX19,
24 network 5G,
25 sample rate 0100, (i.e. 100 Hz)
26 "gps interval" 01 (once per hour~-~- don't change),
27 start mode (always 0),
28 stop mode (always 0),
29 seismometer type (3 = broadband (+/- 20V), 2 = Guralp (+/- 10V), 1 = shortperiod (+/- 5V),
30 and seismometer serial number (4864).
31
32 (% class="box warningmessage" %)
33 (((
34 NOTE: the 2 at the very end is for "RECORD ON RESTART". The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume. (# of blank spaces before this doesn't matter)
35 )))
36
37 === The format for TSAWR loggers is shorter: ===
38
39 (% class="box errormessage" %)
40 (((
41 **NOTE: The format is shorter as they don't have an option to set the "GPS sync". The GPS sync number should be omitted otherwise the formatting will not work!**
42 )))
43
44 SITE15G02500039999 2
45
46 corresponding to site SITE1, 
47
48 network 5G
49 sample rate 0250, (i.e. 250 Hz)
50 start mode (always 0),
51 stop mode (always 0),
52 seismometer type (3 = broadband (+/- 20V), 2 = Guralp (+/- 10V), 1 = shortperiod (+/- 5V),
53 and seismometer serial number (9999).
54
55
56 === The formatting process using the logger: ===
57
58 The process for formatting an SD card within the logger is straightforward. Navigate to the "SD INFORMATION" screen and press ERASE SD CARD. This process may take up to a minute. This will result in erasing all files from the card. Upon starting recording, a new 'seed' will be written containing all the information that the logger has been set with FINISH THIS SECTION
59
60 [[image:original_0747763c-e3dd-4667-b897-833f12c8e0b8_20241219_130604.jpg||height="240" width="298"]]
61
62 Here is a general guideline for how much data you can expect to fit on a card:
63
64 {{{ 64Gb card: 245 days @ 250hz or 610 days @ 100hz
65 32Gb card: 122 days @ 250hz or 305 days @ 100hz
66 }}}
67
68
69 = Logger Menus Overview (and setup) =
70
71 The same menus are used in both the TSAWR and LPR-200 loggers.
72
73 == System Information ==
74
75 This menu provides an overview of the status of the station. Under 'Initialisation Parameters', the status of systems that are initialised upon powering on the logger can be checked.
76
77 This menu also displays the firmware version, battery, external, and solar voltages, and the temperature of the system.
78
79 ==== Upon setup ====
80
81 - Check all Initialisation Parameters are marked as successful.
82
83 - Check that solar voltage is above 10 V, otherwise the station will not last long.
84
85
86 == Live Seismometer Data ==
87
88 This screen displays real time seismometer data for all 3 axes of the seismometer. Initially, a plot of raw data from all 3 channels is shown. By pressing enter, a high pass filter can be applied showing a more useful plot of all channels. The up and down arrows can be used to change between a view of all channels, individual channels, and all channels on different X axes.
89
90 ==== Upon setup ====
91
92 - Check that all 3 channels are present and are producing a signal.
93
94 - Stomp on the ground near where the sensor is installed to check that a signal is being picked up by all 3 channels.
95
96 - Common representations of poor signals include channels sloping smoothly from high values to zero, or sharp corners in the plots.
97
98
99 == GPS Data ==
100
101 This menu displays the status of the stations' GPS connection. The screen lists; UTC time, UTC date, latitude, longitude, altitude, number of satellite connections, and SNR.
102
103 ==== Upon setup ====
104
105 - Check that the station is connected to satellites
106
107
108 == SD Information ==
109
110 The menu lists if the SD card has been initialised, as well as the capacity and space free on the memory card. See "Data card formatting" (above) to see how to format and initialise the inserted SD card.
111
112
113 == System Configuration ==
114
115 This menu is used to set the stations identifiers and parameters.
116
117 First, the sampling rate can be chosen from a set of options (1 Hz, 10 Hz, 25 Hz, 40 Hz, 50 Hz, 100 Hz, 250 Hz, 1000 Hz).
118
119 Next, the station identifier (up to 5 characters) and the network code (2 characters) can be set.
120
121 Record start and stop mode should be left to 'On Request' mode.
122
123 The seismometer model and serial number (up to 10 characters) can be set. Seismometer model options include:
124
125 - Trillium Compact
126
127 - CMG - 3ESP
128
129 - Guralp 40T
130
131 - LE-3D Lite
132
133 - Mark L4C
134
135 - Mark L4
136
137 Additionally, the following settings can be used in place of the seismometer model types: +/- 20V, +/- 10V, +/- 5V
138
139 The record on restart option ensures that if the logger dies and is powered back up whilst in the field (due to battery charging cycles or other causes) that the recording will resume.
140
141 ==== Upon setup ====
142
143 - Set the sampling rate, station identifier, and network code. Set the appropriate seismometer type. Most importantly, ensure the record on restart setting is checked.
144
145 (% class="box errormessage" %)
146 (((
147 NOTE: Ensure the 'RECORD ON RESTART' option is marked with a cross.
148 )))
149
150 (% class="wikigeneratedid" %)
151 [[image:original_2b83fcf5-ee5a-4375-bab1-2d738e4ffe8c_20241219_130646.jpg||height="234" width="289"]]
152
153 == STA/LTA Configuration ==
154
155 The Short Term Average and Long Term Average Configuration screen is no longer used.
156
157
158 = LogFile Conversion Script =
159
160 Both the TSAWR and LPR-200 write logging information as a binary "dat" file which includes GPS time syncs, temperatures, battery power, and position. To convert them into ASCII you can read, use [[THIS PYTHON SCRIPT>>http://auspass.edu.au/field/anusr_log.py]].
161
162
163 = Instrument Response =
164
165 Both the TerraSAWR and LPR-200 use the same ADS1281 analog-to-digital converter chip and are designed to have identical instrument response. Depending on the output sample rate (e.g. 100 Hz, 250 Hz, 500 Hz, 1000 Hz) amplitude response is consistently flat up to ~~100 Hz but phase response can vary above 1 Hz at 100 Hz (or 10 Hz at 250 Hz).
166
167 The user can choose to apply a 2nd stage "sensor gain" by selecting an instrument type in the setup menu. This effectively selects a 10 Vpp (e.g. short period sensors), 20 Vpp, 40 Vpp (most broadband sensors) regime to match the sensor's sensitivity. This has the effect of doubling amplitude from 10v to 20v, or quadrupling from 10v to 40v. If you have set your sensor correctly (and the signal isn't clipped!) you can "correct" this by simply multiplying your data by 0.5 etc.
168
169 Another important thing to note is that the group delay associated with late stage FIR filters is **automatically applied in the logger**, hence there is no need to apply this in the response. These tend to max out at 0.124 seconds for most output sampling rates (0.062 s for 100 Hz).
170
171 Instrument response can be downloaded from IRIS-NRL (v2) if need be, or by downloading the response of an equivalent sensor at AusPass (e.g. get_stations(level='response') )
172
173
174 = ANU TerraSAWR (Gen 3, FW 3.5a, 2017?- current) =
175
176 Text here
177
178 == Sub-paragraph ==
179
180 XXXX
181
182 == Sub-paragraph ==
183
184 xx
185
186 === Sub-sub paragraph ===
187
188 x
189
190
191 = ANU LPR-200 (Gen 2, FW 2.6a/2.7a, 2013 - current) =
192
193 Text here
194
195 == Sub-paragraph ==
196
197 x
198
199 == Sub-paragraph ==
200
201 x
202
203
204 = ANU "ANUSR" (Gen 1, 2003? - 2013?) =
205
206 x
207
208 = References =
209
210 * [[PDF Manual>>http://auspass.edu.au/field/LPR-200_Instruction_Manual.pdf]]
211 )))
212
213 (% class="col-xs-12 col-sm-4" %)
214 (((
215 (% class="box" %)
216 (((
217 **Contents**
218
219 {{toc/}}
220 )))
221
222
223 (% class="box" %)
224 (((
225 = TerraSAWR Specs =
226
227
228 [[image:Terrasawr 1.jpg]]
229
230 [[image:1704865206800-824.jpg]]
231
232 [[image:1704864886951-793.jpg]]
233
234 |=Size (L x W x H)|30 x 25 x12cm
235 |=Weight (with battery)| ~*~*1.9kg (2.5kg)
236 |=Battery|(((
237 Lead-acid 12v, ~~7Ah
238
239 Rechargeable
240 )))
241
242 [[image:terrasawr battery.jpg]]
243 )))
244
245
246 (% class="box" %)
247 (((
248 = LPR-200 Specs =
249
250 [[image:LPR 1.jpg]]
251
252 [[image:LPR 2.jpg]]
253
254 [[image:LPR 3.jpg]]
255
256 |=Size (L x W x H)|42 x 34 x17 cm
257 |=Weight (with battery)|4.9 kg (7.3 kg)
258 |=Battery|(((
259 LiFePo4 140Ah
260
261 Rechargeable
262 )))
263
264 [[image:LPR battery.jpg]]